Main Protocol Performer: Kuba Purpose / Hypothesis Whether pulse can be detected instead of needle depth Equipment and Setup Needle size: 22G Needle tip: Quincke Used sensor: M.4.4.1.22G Punctured material: Candle gel Phantom setup:A fully filled tube embedded inside a angled surface candle gel box. We needed needle tip distances relative to the tube, with different needle depth in the same time. For this purpose, a candle gel box with an… See the full description on the dataset page: https://huggingface.co/datasets/VibroNav/May26_CandleGel_Pump_Experiment.

sourceHugging Facecc-by-nc-4.0updated 3mo agoView on Hugging Face
0likes23downloads
Dataset Card

Main Protocol

Performer: Kuba

Purpose / Hypothesis

Whether pulse can be detected instead of needle depth


Equipment and Setup

  • Needle size: 22G
  • Needle tip: Quincke
  • Used sensor: M.4.4.1.22G
  • Punctured material: Candle gel

Phantom setup: A fully filled tube embedded inside a angled surface candle gel box. We needed needle tip distances relative to the tube, with different needle depth in the same time. For this purpose, a candle gel box with an embedded tube was built.

[image]


Experimental Setup

  • Tube size: 10 × 8 mm
  • Pump speed: 22 rpm
  • Measurements performed at room temperature

Distances

  • Vertical distance from tube center to bottom of box: 3.5 cm
  • Distance from tube center to top surface: 6.3 cm
  • Candle gel surface begins at a radial distance of 6.3 cm from the tube center. In highest point due to angled-surface configruation.

Measurement Points

For each column and each row many recordings were made for different distane from needle tip to tube center.

Each consecutive recording is separated by 5 mm along the radial direction toward the tube center.

The distance to the tube in the Z axis in the recorded signal can be taken from the filename (see Filename Meaning section).

The starting point in the Z axis was different for columns due to angled-surface.

image


Surface Measurement Pattern

Measurements were performed on a grid placed on the candle gel surface.

Grid Layout

  • 6 columns: e, f, g, h, i, j
  • Each column contained 11 rows (10 rows + tube row)

The central row corresponds to the row where the tube is located:

  • 5 rows above the tube
  • 5 rows below the tube

Row Spacing

  • 10 mm spacing between most adjacent rows (1–5, 6–10)
  • 15 mm spacing between:
  • row 5 and tube
  • tube and row 6

[image]


Filename Meaning

Example filename: \ e100.5FoamSpeed-322gquincke120mmM.4.4.1.22G2026-06-11_11.29.04.wav \

Format: \ column-rowdistanceToTubeInCMSmaterialspeedneedle-typesensor-typetimestamp

distanceToTubeInCM/S

  • Measured in the Z axis
  • Independent distance originating from the row

Additional notes:

  • Sensor placed on channel 0
  • Channel 1 not connected

[image]